1、Product introduction
      All medical nickel-titanium compositions of Madis alloy comply with the requirements of ASTM F2063-12 or GB 24627-2009 for surgical implantation. Raw materials must undergo qualified third-party inspection to ensure chemical composition, phase transition temperature, material uniformity and micro-structure requirements. High precision diamond drawing die is used to ensure the diameter and surface finish of finished wire.




2、Composition requirements

Element content,mass%
Ni 54.5-57.0
C ≤0.050(500 ppm maximum)
Co ≤0.050(500 ppm maximum)
Cu ≤0.010(500 ppm maximum)
Cr ≤0.010(500 ppm maximum)
H ≤0.005(50 ppm maximum)
Fe ≤0.050(500 ppm maximum)
Nb ≤0.025(250 ppm maximum)
N+O ≤0.050(500 ppm maximum)
Ti allowance

3、Performance index
Tensile strengthsb ≥1000MPa
Elongationδ% ≥10
Upper Platform Stress ≥400 MPa
Residual Deformation% <0.5
Actual Operation Completion Temperature -30℃~90℃

4、Reference criteria
GB 24627-2009   Nickel-Titanium Shape Memory Alloy Processing Materials for Medical Devices and Surgical Implants  
GB/T 228-2002   Tensile Test Method for Metal Materials at Room Temperature  
YYT 0695-2008   Standard test method for corrosion susceptibility of small implant devices by cyclic potentiodynamic polarization  
ASTM F2082-16   Standard for Measuring Phase Transition Temperature of Nickel-Titanium Shape Memory Alloy by Bending and Free Recovery Method  

5、Supply Specification
5.1、Specifications and tolerances for nickel-titanium alloy wire supply
0.05-0.15mm ±0.004mm
0.15-0.26mm ±0.005mm
0.26-0.35mm ±0.007mm
0.35-1.0mm ±0.01mm
1.0-2.0mm ±0.015mm
2.0-4.0mm ±0.02mm
≥4.0mm Negotiate with customers to determine
It can be decided according to the special needs of customers.
5.2、Supply specifications and tolerances of nickel-titanium alloy tubes
≤1.5 mm ±0.025mm
1.5-6.5 ±0.050mm
6.5-10 ±0.075mm
Note: Oversize Tolerance and Madison Negotiation



6、Notices for Order
According to ASTM F2063-12 or GB 24627-2009, the following information should be provided when inquiring or ordering:
 6.1    AF Temperature in Delivery State  
6.2   Dimensions, such as diameters and tolerances, can be negotiated for more stringent tolerances, and half of the diameter tolerances can be selected for non-roundness tolerances  
6.3   Delivery heat treatment state: straightening superelastic state, cold working state, need follow-up heat treatment  
6.4   Delivery surface state: black oxidized state, light oxidized state (from golden to brown), clean state, bright state, abrasive state  
6.5   Mechanical properties: tensile strength and elongation  
6.6   Order Quantity: Weight, Length or Root Number  

7、Packaging
      The general packaging of Madison wire is I-wheel packaging. If customers have other packaging requirements, they can negotiate and confirm

Suqian Medix Alloy Co., Ltd

Tel

Email

Web

Add

13806169807 Mr. Dai

364969871@qq.com

www.medixalloy.com

No. 26 Haining Road, Shuyang Economic and Technological Development Zone

Feedback

QR code

Scanning Entry